Gx
Series
EN - 42
4. Touch the
icon to start the calibration. In the
R
CAL
field
the resistance of test leads is shown. This
valie will be automatically subtracted by the instrument
at the end of Loop measurement
The instrument displays the
symbol to indicate
the positive outcome of teast leads calibration (
Rcal
<1
) and the screen to the side appears on the
display
5.
Touch the
icon to back to the measurement
main screen. Note the
green icon which
means the successful test leads calibration and
proceed with the measurements described in the
following paragraphs
6. The value of the test leads/Shuko plug resistance
various plugs is maintained by the instrument up to
the reset operation performed by the user (for
example for the insertion of cables with different
lengths). To performs the reset of saved calibration
value, touch the
icon. The screen to the side
appears on the display
7. With open input terminals touch the
icon. The
“
> 1
” indication is shows for a while in the
R
CAL
field
and the “Calibration Reset…” is shown at display.
Touch the
icon to back to the previous screen
(note the “- - -“ indication in the
R
CAL
field) and repeat
the previous steps in order to perform a new
calibration