Gx
Series
EN - 116
40
If the test fails, the screen like the one on the side is
shown on the display. The incorrect value of the
parameter is shown in red and the symbol
indicates
the non-ok result of the measurement
.
Press again the
GO/STOP
key to repeat the test
Press the
SAVE
key or touch the icon
for the
partial saving of the test and to
end the test sequence
.
The instrument shows the message shown in the
followed screen for a few seconds
41 Repeat the sequence again if necessary
42
Act on the three selector switches of the adapter by
setting the followed positions as shown by the
instrument with flashing red / blue indications
:
PP State
13A,20A,32A or 63A
CP State
C
Fault
E
NOTE: Some EVSE stations could not manage this
error condition. In this case, leave the Fault
selector in the PE position to perform the test
Press the icon
to activate the
test on E state
or
press the icon
to exit the test and return to the
initial screen
43
The result of the measurement is shown in the screen
to the side
. The symbol
indicates the correct result
of the measurement.
The symbol
also indicates a
positive result of the measurement but with values
outside the limits present in the Table 4
Press again the
GO/STOP
key to repeat the test
Press the
SAVE
key or touch the icon
to end the
status control test and save the final result in the
instrument's memory (see § 7.1)