Gx
Series
EN - 23
6. Press the
GO/STOP
key on the instrument or the
START
key on the remote lead. The
instrument will start the insulation measurement in sequence before between L-PE
and then between N-PE
CAUTION
During this whole stage, do not disconnect the test leads of the instrument
from the conductor under test. It could remain charged with a dangerous
voltage due to the stray capacitances in the circuit being tested.
7. Regardless of the operating mode selected, at the end of the measurement, the
instrument applies a resistance to the output leads to discharge the capacitances in the
circuit.
8. The measurement result is shown in the screen to the
side. The values of the real test voltage and the
measurement time are present on the display.
The
symbol indicates the positive result of the
measurement.
Press the
SAVE
button or touch the
icon to save
the measurement (see § 7.1)
9. At the end of the test, if the value of the measured
resistance in one or both test should be lower than the
set limit, the screen to the side is shown on the display.
The value is shown in red and the
symbol indicates
the negative result of the measurement.
Press the
SAVE
button or touch the
icon to save
the measurement (see § 7.1)