Gx
Series
EN - 17
Anomalous situations
6.1.1.
1. In AUTO or "
" modes if the instrument detects a
resistance lower than the set limit value but for which is it
not able to generate a current of 200mA, the screen to
the side is displayed.
The
symbol is shown on the display and the values
of the real test current are indicated in red.
2. If in
mode the instrument detects a resistance
higher than 5
at its terminals, it resets the offset value
and displays a screen like the one to the side. The
icon is shown on the display to indicate the
calibration reset value (i.e. performing the operation with
open terminals).
3. In case the instrument detects a calibration resert
(performing the operation with open terminals), it gives
out a long sound and displays a screen like the one to
the side. The
icon is shown on the display to
indicate the calibration reset value.
4. If the instrument detects a calibrated resistance higher
than measured resistance (e.g. by using test cables
different from the supplied ones) at its terminals, it gives
out a long sound and displays a screen like the one to
the side. Perform a calibration reset with open terminals
and start a new calibration.
5. If the instrument detects a voltage value higher than 10V
at its terminals, it does not carry out the test, gives out a
long sound and the screen reported here to the side is
displayed.