Gx
Series
EN - 14
1.
Touch the
icon. The screen to the side appears on
the display. The instrument automatically carries out the
test for the presence of voltage between the inputs
(shown on the display) and blocks the test in case of
voltage higher than 10V. Touch the “AUTO” icon to set
the measuring mode. The following screen appears on
the display:
2. Move the slide bar reference in the positions "AUTO"
(Automatic mode), " " (Timer mode) or “ ” (Multimeter
mode). Confirm the choice by going back to the previous
screen. If Timer mode is selected, the following screen is
shown:
3.
Touch the
icon to zero the value in the Timer field
and use the virtual keyboard to set the value in seconds
between
1s
and
15s
. Confirm the choice by going back to
the initial measurement screen.
4. Touch the icon "R
xx
" to set the maximum limit value of
the resistance on which the instrument makes the
comparison with the measured value. The screen to the
side appears on the display.
Touch the
icon to zero the value in the “R
” field.
Use the virtual keyboard to set the value between
0.01
and
9.99
.
Confirm the choice by going back to the initial
measurement screen. Note the presence of the set limit
value.
5. Perform, if necessary, the compensation of the measuring leads resistance by
connecting the cables or the remote lead as shown in Fig. 7.
Fig. 7: Compensation of single cables and remote lead resistance