617
Error Messages
One or more self tests failed. Error code: XXXX
An instrument failure was detected when the Test Set was turned on. (For example, having
a stuck front-panel key during turn on.) The numbered error message corresponds to a
binary-weighted group of errors listed in the *TST Common Command description in the
Programmer’s Guide.
Change Ref Level, Input Port or Attenuator (if using “Hold”).
The RF signal level is either too great or too small for the current input port, attenuator
setting, or both. This error often occurs when trying to make a low-level measurement
using the RF IN/OUT port with the Spectrum Analyzer. Make the indicated changes until
this message is no longer displayed.
Change RF Gen Amplitude, Output Port or Atten Hold (if on).
This message appears when the RF Generator’s
Amplitude
field is set too high when
using the RF IN/OUT port or when adjusting the amplitude with the
Atten Hold
field
set to
On
.
The RF IN/OUT port has a lower maximum output level than the DUPLEX OUT port.
Use the DUPLEX OUT port, or reduce the RF Generator’s level.
If
Atten Hold
is
On
, you may be adjusting the amplitude outside of the allowed range.
Change the amplitude
Direct latch write occurred. Cycle power when done servicing.
The SERVICE screen was accessed and one or more internal latch settings were changed.
Turn the instrument off and back on to reset the latches. (This condition can occur during
periodic calibration.)
Summary of Contents for 8920A
Page 14: ...Contents 14 2 Configuring Your Test Set General Operating Information 50 ...
Page 17: ...Contents 17 5 AF Analyzer Screen Block Diagram 100 ...
Page 19: ...Contents 19 7 Configure Screen Field Descriptions 254 ...
Page 21: ...Contents 21 9 Duplex Test Screen Block Diagram 340 Field Descriptions 341 ...
Page 23: ...Contents 23 11 Help Screen Field Descriptions 436 ...
Page 24: ...Contents 24 12 I O Configure Screen Field Descriptions 438 ...
Page 25: ...Contents 25 13 Message Screen Field Descriptions 444 ...
Page 26: ...Contents 26 14 Oscilloscope Screen Field Descriptions 446 Using the Oscilloscope 452 ...
Page 27: ...Contents 27 15 Print Configure Screen Field Descriptions 454 ...
Page 29: ...Contents 29 17 RF Analyzer Screen Block Diagram 472 Field Descriptions 473 ...
Page 30: ...Contents 30 18 RF Generator Screen Block Diagram 484 Field Descriptions 485 ...
Page 31: ...Contents 31 19 RX Test Screen Block Diagram 494 Field Descriptions 495 ...
Page 32: ...Contents 32 20 Service Screen Field Descriptions 502 ...
Page 35: ...Contents 35 23 TX Test Screen Block Diagram 550 Field Descriptions 551 ...
Page 38: ...Contents 38 Index 619 ...
Page 39: ...39 1 Get Started ...
Page 99: ...99 5 AF Analyzer Screen ...
Page 113: ...113 6 Call Processing Subsystem ...
Page 268: ...268 S HP892XX USRGUIDE manual confscrn fb Chapter 7 Configure Screen Field Descriptions ...
Page 269: ...269 8 Signaling Decoder Screen ...
Page 339: ...339 9 Duplex Test Screen ...
Page 353: ...353 10 Signaling Encoder AF Generator 2 ...
Page 435: ...435 11 Help Screen ...
Page 437: ...437 12 I O Configure Screen ...
Page 443: ...443 13 Message Screen ...
Page 445: ...445 14 Oscilloscope Screen ...
Page 471: ...471 17 RF Analyzer Screen ...
Page 482: ...482 S HP892XX USRGUIDE manual rfanscrn fb Chapter 17 RF Analyzer Screen Field Descriptions ...
Page 483: ...483 18 RF Generator Screen ...
Page 492: ...492 S HP892XX USRGUIDE manual rfgscrn fb Chapter 18 RF Generator Screen Field Descriptions ...
Page 493: ...493 19 RX Test Screen ...
Page 504: ...504 S HP892XX USRGUIDE manual servscrn fb Chapter 20 Service Screen Field Descriptions ...
Page 519: ...519 22 Tests Screen ...
Page 548: ...548 S HP892XX USRGUIDE manual testscrn fb Chapter 22 Tests Screen ROM Programs ...
Page 549: ...549 23 TX Test Screen ...
Page 561: ...561 24 Connector Key and Knob Descriptions ...
Page 585: ...585 25 Modifications Accessories Manuals Support ...
Page 605: ...605 Error Messages ...
Page 618: ...618 S HP892XX USRGUIDE manual errormsg fb Error Messages ...