8 Troubleshooting
100
COD
HIGH
Unsuitable upper limit of
setting value.
Review the upper limit of the
setting value.
"5.5.6 Setting of MAX Alarm
Value for COD and TURB
(turbidity)" (page 53)
Abnormal upper limit of
sample water.
Check the sample water.
Incompatible calibration
solution.
Perform zero and span
calibration.
Contamination of measuring
cell.
Clean the measuring cell.
Replace the wiper blade
rubber when it is
deteriorated.
"7.3.1 Cleaning of Measuring
Cell" (page 84)
"7.3.2 Inspection and
Replacement of Wiper Blade
Rubber" (page 86)
LOW
Unsuitable lower limit of
setting value.
Review the lower limit of the
setting value
"5.5.7 Setting of MIN Alarm
Value for COD and TURB
(turbidity)" (page 54)
Abnormal upper limit of
sample water.
Check the sample water.
Incompatible calibration
solution.
Perform zero and span
calibration.
Contamination of measuring
cell.
Clean the measuring cell.
Replace the wiper blade
rubber when it is
deteriorated.
"7.3.1 Cleaning of Measuring
Cell" (page 84)
"7.3.2 Inspection and
Replacement of Wiper Blade
Rubber" (page 86)
Item
Alarm
Causes
Measures
Refer