6 External Input and Output
73
6.2
Analog Output
This unit outputs the measurement value by analog current output when outputting to the
outside.
The analog output contains three outputs, and outputs by selecting three components from
the following five components.
The output value is output in conjunction with the full scale of each measurement value.
Measurement item and analog output range: example of 4 to 20 mA output
Measurement item and analog output range: example of 0 to 16 mA output
Allocations for measurement analog output terminals are as follows.
Specifications for analog current output are as follows.
Reference
"5.4.1 Signal Allocation" (page 37)
Measurement item
Range (example)
4 mA output
20 mA output
UV
1 Abs
0 Abs
1 Abs
VIS
1 Abs
0 Abs
1 Abs
UV-VIS
1 Abs
0 Abs
1 Abs
COD
100 mg/L
0 mg/L
100 mg/L
TURB (turbidity)
100 mg/L
0 mg/L
100 mg/L
Measurement item
Range (example)
0 mA output
16 mA output
UV
1 Abs
0 Abs
1 Abs
VIS
1 Abs
0 Abs
1 Abs
UV-VIS
1 Abs
0 Abs
1 Abs
COD
100 mg/L
0 mg/L
100 mg/L
TURB (turbidity)
100 mg/L
0 mg/L
100 mg/L
Terminal No.
Output No.
Allocation
17
18
A-1(+)
A-1(
)
(Arbitrary setting)
5
6
A-2(+)
A-2(
)
(Arbitrary setting)
19
20
A-3(+)
A-3(
)
(Arbitrary setting)
Signal type
Input and output
circuit
Specifications
Analog signal output
DC 0 mA to 16 mA/4 mA to 20 mA
Current signal output
Insulated output (common to COM)
Load resistance: Max. 600
Ω