4 Calibration
33
4.
Pour the residual TURB span solution into the cleaning bottle.
5.
Take the analyzer unit from the calibration tank, and clean the cell unit using the
TURB span solution in the cleaning bottle.
Note
Do not spill any solution on the base of cell rotary shaft (V-ring seal part).
When removing the analyzer unit, make sure not to lose the wing nuts.
As the UV lamp is illuminated, work with the attached light shielding glasses. The UV lamp should
not be viewed directly as your eyes may be damaged.
6.
After cleaning the cell unit with the TURB span solution, dip in the calibration tank.
7.
Press the button for [SPAN VALUE] on the TURB. CAL. screen.
The screen changes to the TURB. SPAN screen.
Residual TURB span solution
Cleaning bottle
Clean the cell unit using the TURB span solution
in the cleaning bottle.
Allow the solution to adhere by shaking from side to side for several times.
[x.xx]