5 Functions
60
5.8
Data Check
Items for the data check are as follows.
Data Screen Display
1.
Press the [DATA] on the MEAS. screen.
The DATA screen appears.
Items for data check
Description
LOG DATA check
Checks the previous data readings numerically. Approximately11 days
can be saved for the value of 1 minute and approximately 13 months can
be saved for the value of 1 hour.
LOG DATA deletion
Deletes all of the log data.
GRAPH display
Checks the previous readings of data using a graph.
CAL. REPORT check
Checks up to 20 history records by measurement component and zero/
span.
CAL. REPORT deletion
Deletes all of the calibration report.
ALARM REPORT check
Checks up to 511 alarm history records.
ALARM REPORT deletion
Deletes all of the alarm history.
CF CARD transfer
Transfers the data to the CF card.
CF CARD initialization
Initializes the CF card.
[DATA]
MEAS. screen