5 Functions
41
1.
Press the output item button to be set on the OUTPUT SETTING screen.
MAINT. OUTPUT
WARM-UP OUTPUT
ALARM OUTPUT
The screens for the output of each item appear as follows.
The screen shows an example for MAINT. OUTPUT.
2.
Select the setting item, and press [SET]
The screen returns back to the OUTPUT SETTING screen.
Output settings during operation are complete.
★
When setting to PRESET:
1.
Press the [PRESET] button on the lower right of screen.
The PRESET screen for each item appears.
2.
Enter the preset value with the numeric keypad, and press [SET],
Input range: 0.0 mA to 21.0 mA
Note
When it is preset, the log data also becomes the preset value. The value corresponds to the following
calculating formula.
When the analog output level is 4 mA to 20 mA:
Full scale of measurement value × (preset value
4) / 16
When the analog output level is 0 mA to 16 mA:
Full scale of measurement value × preset value / 16
Tip
If the setting is changed while holding the readings, the setting change is enabled after the hold is
released.
[HOLD]
/[DIRECT]
/[PRESET xxmA]
Setting item
Description
Initial setting
HOLD
Outputs the adjacent readings while holding.
If there are no adjacent readings when the power is ON,
the latest readings before the power OFF are held.
-
DIRECT
Outputs the measured readings directly.
-
PRESET
Outputs the preset values.
In this case, the log data is the reading equivalent to the
setting output value.
During maintenance and alarm: 21.0 mA
During warm-up: 4.0 mA
[PRESET]