8 Troubleshooting
99
UV
ZERO
CAL.
Incompatible calibration
solution.
Make sure to use distilled water
for the calibration solution.
Contamination of measuring
cell.
Clean the measuring cell.
Replace the wiper blade
rubber when it is
deteriorated.
"7.3.1 Cleaning of Measuring
Cell" (page 84)
"7.3.2 Inspection and
Replacement of Wiper Blade
Rubber" (page 86)
Failure of parts in the
analyzer unit.
Replace the faulty parts.
Service call
SPAN
CAL.
Incompatible calibration
solution.
Make sure to use the exclusive
calibration solution, and distilled
water as the dilution water.
"4.4.3 Common Span
Calibration" (page 24)
Contamination of measuring
cell.
Clean the measuring cell.
Replace the wiper blade
rubber when it is
deteriorated.
"7.3.1 Cleaning of Measuring
Cell" (page 84)
"7.3.2 Inspection and
Replacement of Wiper Blade
Rubber" (page 86)
Failure of analyzer unit parts. Replace the faulty parts.
Service call
VIS
ZERO
CAL.
Incompatible calibration
solution.
Make sure to use distilled water
for the calibration solution.
Contamination of measuring
cell.
Clean the measuring cell.
Replace the wiper blade
rubber when it is
deteriorated.
"7.3.1 Cleaning of Measuring
Cell" (page 84)
"7.3.2 Inspection and
Replacement of Wiper Blade
Rubber" (page 86)
Failure of analyzer unit parts. Replace the faulty parts.
Service call
SPAN
CAL.
Incompatible calibration
solution.
Make sure to use the exclusive
calibration solution, and distilled
water as the dilution water.
"4.4.3 Common Span
Calibration" (page 24)
Contamination of measuring
cell.
Clean the measuring cell.
Replace the wiper blade
rubber when it is
deteriorated.
"7.3.1 Cleaning of Measuring
Cell" (page 84)
"7.3.2 Inspection and
Replacement of Wiper Blade
Rubber" (page 86)
Failure of analyzer unit parts. Replace the faulty parts.
Service call
Item
Alarm
Causes
Measures
Refer