5 Functions
42
5.4.4
Output Condition Setting
Here enables the output conditions setting (such as the alarm discrimination time and the
minus display).
For normal usage, default setting can be used (no additional setting is required).
1.
Press [OUTPUT CONDITION SETTING] on the SETTING 1/2 screen.
The OUTPUT CONDITION SETTING screen appears.
2.
Press the button for ALARM DISCRIMI. TIME to change the discrimination time of
the alarm output, and enter the value of setting range. Press the button for MINUS
DISPLAY to activate/deactivate the minus display.
The screen returns back to the OUTPUT CONDITION SETTING screen.
The output condition setting is complete.
[MAINT.]
MEAS.
[OUTPUT CONDITION SETTING]
[SETTING]
Refer to "5.4.2 Signal Input Setting" (page 39).
OUTPUT CONDITION SETTING
MAINTENANCE
SETTING 1/2
screen
screen
screen
screen
[OUTPUT CONDITION
SETTING]
Setting item
Description
Initial setting
ALARM DIS-
CRIMI. TIME
Sets the alarm delay time to ignore the spike alarm.
When the time exceeds the ALARM DISCRIME. TIME, the alarm is
triggered and sounds continuously.
(Set to "0" minute if you want to occur the alarm immediately.)
Setting range: 0 to 99 minutes
5min
MINUS DIS-
PLAY
Sets whether to display the measurement value.
Setting range: OFF or ON
OFF