5 Functions
50
5.5.4
Setting of Turbidity Correction Factor
The turbidity correction factor "
" used for the calculation of UV-
VIS is set here.
1.
Press the button for TURB. COR. FAC. on the VIS MEAS. SETTING screen.
The VIS TURB. CORRECTION FACTOR screen appears.
2.
Enter the numeric turbidity correction factor "
" and press [SET].
The screen returns back to the VIS MEAS. SETTING screen.
Unit setting for turbidity correction factor is complete.
Note
Enter the turbidity correction factor "
" when the sample water contains much turbidity to cause a
poor correlation between the UV value and COD value by manual analysis.
[MAINT.]
MEAS.
[VIS]
[MEAS.SETTING]
Refer to "
Setting Screen Display for each Measuring Component" (page 46).
VIS MEAS. SETTING
MAINTENANCE
MEAS. SETTING
screen
screen
screen
screen
TURB. COR. FAC.
(
)