5 Functions
40
Note
Only the time within ±20 minutes of current set time can be input to set the correct clock time. If the
time out of this range is input, the setting becomes invalid.
5.4.3
Output Setting
Here enables the output parameter setting.
1.
Press [OUTPUT SETTING] on the SETTING 1/2 screen.
The OUTPUT SETTING screen appears.
2.
Press the button for ANALOG OUTPUT LEVEL.
The ANALOG OUTPUT LEVEL screen appears.
3.
Select the setting item from [0 - 16] or [4 - 20] and press [SET].
The screen returns back to the OUTPUT SETTING screen.
Setting for the analog output level is complete.
Output settings during operation
Here describes the following three output setting methods.
MAINT. OUTPUT
WARM-UP OUTPUT
ALARM OUTPUT
[MAINT.]
MEAS.
[OUTPUT SETTING]
[SETTING]
Refer to "5.3 Maintenance Screen Display" (page 36).
OUTPUT SETTING screen
MAINTENANCE
SETTING 1/2
screen
screen
screen
Setting item
Description
Initial setting
ANALOG OUTPUT
LEVEL
Sets the analog output level.
Setting range: 0 mA to 16 mA, 4 mA to 20 mA
4 mA to 20 mA
MAINT. OUTPUT
Sets the measurement output during calibration/maintenance to
output by one of the settings below.
Setting range: HOLD, DIRECT, PRESET (0 mA to 21 mA)
HOLD
WARM-UP OUTPUT
Sets the measurement output during warm-up to output by one of
the settings below.
Setting range: HOLD, DIRECT, PRESET (0 mA to 21 mA)
PRESET
4.0 mA
ALARM OUTPUT
Sets the measurement output during an alarm to output by one of
the settings below.
Setting range: HOLD, DIRECT, PRESET (0 mA to 21 mA)
HOLD
[MAINT.]
MEAS.
[OUTPUT SETTING]
[SETTING]
Refer to "5.3 Maintenance Screen Display" (page 36).
OUTPUT SETTING screen
MAINTENANCE
SETTING 1/2
screen
screen
screen