5.5.6
Setting of MAX Alarm Value for COD and TURB (turbidity)
Here describes how to set the upper limits of the COD and TURB alarm values.
1.
Press the button of the measuring component on the COD or TURB MEAS.
SETTING screen.
The MEAS. SETTING screen for each measuring component appears.
2.
Press the button for MAX ALARM VALUE.
The MAXIMUM ALARM VALUE screen for each measuring component appears.
The screen shows an example of MAX ALARM VALUE for COD.
3.
Enter the upper limit value with the numeric keypad, and press [SET].
Input range:
9999.9 to 9999.9
The described above are provided the sample when the decimal place is set to 1 digit.
Tip
[CLEAR]
:
Clears all input numeric values.
[BACK]
:
Clears the rightmost number.
The screen returns back to the MEAS. SETTING screen for each measuring component.