![horiba OPSA-150 Instruction Manual Download Page 68](http://html1.mh-extra.com/html/horiba/opsa-150/opsa-150_instruction-manual_2162276068.webp)
5 Functions
56
5.6
Maintenance - Action
This function performs operations using the operating buttons.
5.6.1
Action
Here describes the execution methods for wiper motor and UV lamp.
1.
Press the [ACTION] button on the MAINTENANCE screen.
The ACTION screen appears.
2.
Every time the item to be activated is pressed, the value switches between ON and
OFF.
The action is activated. Check whether the value has been activated normally.
3.
Press [CLOSE] to return back to the MAINTENANCE screen.
Press [CLOSE] to return back to the ON condition automatically.
Note
When checking the action, do not touch the cell part, and do not directly look at the UV lamp; it is
dangerous.
When performing the action, the motor and lamp are stopped and abnormal measurement values
may be recorded.
Action
Description
WIPER MOTOR
Checks the ON/OFF operation of the rotary cell wiper motor.
UV LAMP
Checks the ON/OFF operation of the UV lamp for light.
[MAINT.]
MEAS. screen
[ACTION]
Refer to "5.3 Maintenance Screen Display" (page 36).
MAINTENANCE screen
ACTION screen
[ACTION]