5 Functions
51
5.5.5
Setting of COD Conversion Factor and TURB Meas. Correction Factor
Here describes how to set the COD conversion factor from UV-
VIS and TURB Meas.
correction factor.
Reference
"9.5 How to Calculate the COD Conversion Factor" (page 106)
1.
Press the button for y = a + bx on the COD or TURB MEAS. SETTING screen.
In case of COD, up to five pairs of the COD conversion factors “a” and “b” are available.
The screen for selecting the factor appears.
In case of TURB, the screen for selecting the factor “a” or “b” appears.
The screen shows an example for COD.
2.
Select the pair to set the factor, and press [SETTING].
The screen for selecting the factor “a” or “b” appears.
3.
Press [a=x.x].
The screen for setting the factor “a” appears.
[MAINT.]
MEAS.
[MEAS. SETTING]
Refer to "
Setting Screen Display for each Measuring Component" (page 46)
COD MEAS. SETTING
MAINTENANCE
MEAS. SETTING
TURB MEAS. SETTING
[TURB]
[COD]
screen
screen
screen
screen
screen
[a=x.x b=x.xxx]
【
SETTING
】