5.8.1
Log Data Check
Displays the measurement value of 1 minute and 1 hour memorized by the measurement.
1.
Press the [LOG DATA] on the DATA screen.
The LOG DATA screen appears.
The time of log data and readings of each measurement component are displayed.
(MIN): data when the number of seconds is "00". Maximum data: approximately 11 days
(HOUR): data when the number of minutes is "00". Maximum data: approximately 13
months
The following marks are displayed for the data during the action.
* If multiple actions are overlapping, they are displayed in numerical order.
Tip
[DELETE]
:
Deletes the all of the log data.
[
][
]
:
Switches to the previous and next data.
[CHANGE]
:
Switches between (MIN) and (HOUR) LOG DATA.
[CALL]
:
The data at a specified time can be viewed.
View the data at a specified time
1.
Press [CALL].
The CALL TIME (MIN/HOUR) screen appears.
2.
Select the hour to be changed and set the number by using [
] and [
]. After
entering the number, press [SET].
The LOG DATA screen with the specified date and time appears.
[DATA]
MEAS. screen
Refer to "
Data Screen Display" (page 60)
DATA screen
[LOG DATA]
During warm-up
During alarm
During maintenance
①
②
③
[CALL]