Operation
31
F.dt
FAiL output delay time
setting
Specify the time in seconds from when a FAiL
occurs until a contact is output.
Curr
General transmission output setting
C.HoLd
Transmission hold plan in
general
Specify the transmission holding status in
general in the output hold mode.
HoLd, PrES HoLd
b.out
Burnout hold setting for
general transmission
output
Specify the burnout current of the transmission
output when an error occurs during
measurement.
Note:
A burnout current is output in the meas mode
only.
non
Cur2.t
Transmission output 2
range (fixed/automatic
changeover) setting
rng
Cur1
C.rng
Transmission output 1 full-
scale range (fixed range)
setting
100.0
rng.0
Zero point value setting for
transmission output 1
Specify the value at the zero point (4 mA)
of transmission output 1.
Note:
The item of sub display depends on which
ever “10.00”
*1
0.00
*2 0.0 to 110.0 0.0
*3 0 to 2200
0
rng.S
Span point value setting for
transmission output 1
Specify the value at the span point (20 mA)
of transmission output 1.
Note:
The item of sub display depends on which
ever “10.00”
*1
10.00
*2 0.0 to 110.0 100.0
*3 0 to 2200
1000
PrES
Preset value setting for
transmission output 1
Input the preset value for current holding.
in “C.rng” of Cur1 is selected.
*1
10.00
*2 0.0 to 110.0 100.0
*3 0 to 2200
1000
Cur2.A to
Cur2.d
Transmission output 2.A to 2.d setting
Tip
The other transmission output 2.b, 2.c, 2.d can be set in the same manner.
The transmission output 2 provides maximum four output ranges which are selected
by external inputs or automatic range selection function. The zero and span point of
the transmission range can be set any within the measuring range.
C.rng
Transmission output 2.A
value (full-scale range)
setting
Cur2.A
1000
Cur2.b/2.c
100.0
Cur2.d
10.00
Display
Item
Description
Setting
range
Initial
value
Refer
to page