Operation
30
r1 to r3
Contact output setting: R1 to R3
Tip
The other relays can be set in the same manner.
Reference
Refer to “ Control targets and their available options ” (page 48) for the options available
for the selected control target.
1.tyPE
Contact output type setting Specify the contact output type (Control target).
non
1.HorL
Specify whether to perform control using the
upper limit operation or lower limit operation.
Note:
This is displayed when “1.tyPE” is set to “AL”.
Hi
1.rng
Selection of decimal point
position (range) for alarm
setting value
Select the optimal position of the decimal point
(range) for the alarm settings.
Note:
This is displayed when “1.tyPE” is set to “AL”.
100.0
1.SEt
*1
10.00
*2 0.0 to 100.0 2.000
*3 0 to 2000
2000
1.dt
Contact output delay time
setting
Specify the delay time in seconds from when
an alarm or hold state occurs until a contact is
output.
Note:
This is displayed when “1.tyPE” is set to “AL” or
“HoLd”.
1.CLn.t
Output type setting for
cleaning
Specify the cleaning output type.
Note:
This is displayed when “1.tyPE” is set to “CLn”.
CLn
1.CAL.t
Output type setting for
calibration
Specify the calibration contact type.
Note:
This is displayed when “1.tyPE” is set to “CAL”.
CAL
FAiL
Error alarm (FAiL) output setting: RF
F.SC
Setting of FAiL output
when abnormality is
detected during self-check
Specify whether or not to output a FAiL when
an abnormality is detected during self-check.
yES
F.oF
Setting of FAiL output
when measured value
exceeds full-scale range
Specify whether or not to output a FAiL when
the measured value exceeds the full-scale
range.
no
F.CAL.E
Setting of FAiL output
when a calibration error
occurs (Calibration error is
regarded as “FAiL”)
no
Display
Item
Description
Setting
range
Initial
value
Refer