Troubleshooting
68
Measurements exceeding the measurement range
If the measured value of turbidity is higher than the maximum value of the display, which is
determined by the position of the decimal point, the measured value will start blinking.
Take remedial action as instructed below.
Symptom
Possible cause
Remedial action
Blinking display
(outside the
measurement
range)
Concentration of the
sample water is high
If the measured value of zero water is normal, the measurement is outside
the range. No remedial action is possible.
Concentration of the
sample water is low
For a measurement of water that is close to transparent, it is possible that
a negative value is obtained. Perform zero calibration again.
If you immerse it in the sample water of a high concentration and issue a
negative value command, contact us.