Measurement
52
Calibration start condition
The automatic zero calibration is performed according to the following conditions.
Automatic zero calibration operation conditions
Calibration and associated operations
CAL Calibration start type
Calibration
cycle
During
calibration:
Waiting: -
External
contact input
ON:
OFF: -
Calibration
operation
*1
ON:
OFF: -
CAL
Contact
output
*2
ON:
OFF: -
HoLd
Contact
output
*3
ON:
OFF: -
Transmission
output
yES
nor
Started only by
internal timer
4
-
-
-
-
-
-
-
-
-
Measurement
value
or
Started by both
internal timer and
external
4
-
-
-
-
-
-
-
Measurement
value
And
Started by internal
timer when external
4
-
-
-
-
Measurement
value
-
-
-
-
-
-
-
-
-
trg
Started by external
trigger
No setting
ON for 2 s
4
-
-
-
-
Measurement
value
no
No setting
No setting
-
-
4
-
-
-
-
Measurement
value
1: When the “Automatic zero calibration setting” setup menu is set to “yES”, the calibration operation is performed
according to the specified method (
). Zero calibration is not performed when it is to be started during cleaning.
2: The “CAL” contact is output for the specified hold time during automatic zero calibration or after calibration (
).
3: There are two types of “HoLd” contact output: 1. “HoLd” contact is output for the specified hold time during the
automatic zero calibration or after the calibration (
), and 2.”HoLd” contact is output when the external contact
input is ON (
).
4: The transmission output is issued according to the settings in the “Transmission output hold setting” setup menu.