Operation
29
SEnSor
tyPE
Turbidity standard
substance (type) selection
PSL
Unit
Selection of turbidity unit or
SS unit (mg/L) for Kaolin
For Kaolin, turbidity or mg/L can be further
selected for the unit of measurement.
Note:
This is displayed when “tyPE” is set to “SiO”.
PtnA
C.CoEF Coefficient input
You can adjust the measure value to the
analyzed value with this coefficient with Kaolin
standard.
1.00
diSP
d.rng
Selection of automatic
decimal point shift or fixed
decimal point
Select either the fixed decimal point or
automatic decimal point shift.
Auto
S.diSP
Sub display area display
setting
non
rng.C
Setting of a hidden area
outside of measurement
range (range cut)
yES
CAL
Automatic zero calibration setting
CAL
Automatic zero calibration
setting
no
tyPE
Setting of automatic zero
calibration start method
or
CyCLE
Automatic zero calibration
cycle setting
Specify the calibration cycle (in hours) when
the automatic zero calibration function is
available.
1 to 999 (h) 168
HoLd.t
Transmission hold time
setting after completion of
automatic zero calibration
60
StAb.L
Stability level setting during
automatic calibration
0.01 to 0.05 0.01
Stab.t
Stability duration setting
during automatic
calibration
30
CAL.E
Display setting of
automatic calibration
attempt error E-43
Specify whether or not to display an automatic
calibration attempt error (E-43).
no
Display
Item
Description
Setting
range
Initial
value
Refer
to page