Measurement
42
Contact output setting: R1 to R3
Tip
The other relays can be set in the same manner.
Reference
Refer to “ Control targets and their available options ” (page 48) for the options available for
the selected control target.
Contact output
type setting
Specify the contact output type (Control target).
non: Stops the function and no contact is output
AL: Operates the alarm contact point
HoLd: Issues an output in a transmission output hold state
CLn: Outputs a trigger at the end of cleaning or outputs during
cleaning
CAL: Outputs a trigger at the end of the calibration or outputs during
calibration or outputs when a calibration error occurs
non, AL,
HoLd,
CLn,
CAL
Control
direction
setting
Specify whether to perform control using the upper limit operation or
lower limit operation.
Hi: Relay 1 is activated when a value is over the alarm level.
Lo: Relay 1 is activated when a value is under the alarm level.
Note:
This is displayed when “1.tyPE” is set to “AL”.
Hi, Lo
Selection of
decimal point
position
(range) for
alarm setting
value
Select the optimal position of the decimal point (range) for the alarm
settings.
Note:
This is displayed when “1.tyPE” is set to “AL”.
10.00,
100.0,
1000
Alarm value
setting
Specify the alarm value.
The default value is reset to 2.000 or 10.00 according to the selected
range.
Note:
This is displayed when “1.tyPE” is set to “AL”.
The item of sub display depends on which ever “10.00”
*1
or “100.0”
*2
or “1000”
*3
in “1.rng” is selected.
*1
0.00 to
10.00
*2
0.0 to
100.0
*3
0 to 2000
Contact output
delay time
setting
Specify the delay time in seconds from when an alarm or hold state
occurs until a contact is output.
If the alarm is cancelled, the contact will be immediately cancelled
without delay.
Note:
This is displayed when “1.tyPE” is set to “AL” or “HoLd”.
0 to 600
(s)
Output type
setting for
cleaning
Specify the cleaning output type.
CLn: Issues an output in a hold state during and after cleaning
trg: Outputs the contact for 5 seconds starting 2 seconds after the
completion of cleaning
Note:
This is displayed when “1.tyPE” is set to “CLn”.
CLn, trg
Display
Item
Description
Setting
range