![horiba HU-200TB-W SS-120-W Instruction Manual Download Page 51](http://html1.mh-extra.com/html/horiba/hu-200tb-w-ss-120-w/hu-200tb-w-ss-120-w_instruction-manual_2162226051.webp)
Measurement
39
List of items displayed for the setting menu
Display
Item
Description
Setting
range
Parameter setting
Moving
average count
(damping
factor) setting
Increase dF to reduce variations in the measured data:
Decrease dF to speed up the response.
It is possible to change the damping factor in the meas mode.
1 to 120
(times)
Auto return
setting
This setting returns the system from the hold mode to the meas mode
automatically.
yES: Enables auto return
The mode is automatically returned when the specified time has
elapsed after the last key operation.
no: Disables auto return
As the mode is not automatically returned, it is necessary to return to
the meas mode manually.
yES, no
Auto return
time setting
The hold mode will be back to the meas mode automatically when the set
time is passed from the last key touch.
In the hold mode, a transmission signal is in a hold state. To cancel the
hold state, it is necessary to return to the meas mode.
Note:
This is displayed when “A.rEt” is set to “yES”.
1 to 24
(h)
Transmission
hold
cancellation
time setting
This setting allows you to maintain the transmission hold state until the
measured value is stabilized after the hold mode has been switched to
the meas mode. Set the time (in seconds) until the maintained
transmission hold state is cancelled.
2 to 30
(s)
Sensor setting
Turbidity
standard
substance
(type) selection
Select the turbidity standard substance from among Kaolin (SiO),
Formazin (For), and Polystyrene Latex (PSL).
SiO, For,
PSL
Selection of
turbidity unit or
SS unit (mg/L)
for Kaolin
For Kaolin, turbidity or mg/L can be further selected for the unit of
measurement.
Pattern A: TURB (turbidity)
Pattern b: mg/L (SS)
Note:
This is displayed when “tyPE” is set to “SiO”.
PtnA,
Ptnb
Coefficient
input
You can adjust the measure value to the analyzed value with this
coefficient with Kaolin standard.
Note:
This is displayed when “tyPE” is “SiO” and the “Unit” is “Ptnb”.
0.01 to
300.00