Measurement
54
List of items displayed for the cal menu
Display
Calibration
Type
Item
Description
Setting
range
Initial
value
Manual zero calibration
Note
When an automatic zero calibration function (option) is provided, specify “ Automatic zero
calibration setting ” (page 41).
Reference
Refer to “ Zero calibration ” (page 59).
Zero
Manual
zero
calibration
The current measured value is blinking. The value
displayed when ENT is pressed will be set to zero.
Pour zero water into the detector. At this point, if
contamination in the accumulated water will affect the
calibration performance, keep supplying the zero water and
wait until the indication is stabilized.
Once the indication is stabilized, press ENT at the lowest
value possible, and then complete the zero calibration.
-
-
Span calibration
Note
Specify the items of “Turbidity standard substance (type) selection” in advance.
Reference
Refer to “ Span calibration ” (page 61).
Span
Span
calibration
Adjust the current value to the turbidity of the standard
solution.
The measured value of the span calibration liquid is
displayed, and a coefficient is displayed in the sub display
area.
Press
/
to calibrate span and press ENT to finish span
calibration.
A measured value changes in proportion to the coefficient.
Adjust the coefficient so that the turbidity is equal to that of
the turbidity standard substance that was poured into the
detector.
Or, adjust the coefficient so that the value is equal to that of
the already tested span bottle.
The indication is stabilized within one minute after the span
calibration liquid has been poured into the detector. Press
ENT sometime between one to three minutes later.
-
1.0000
Data shift by adjust
Manual
shift
Adjust
You can shift the measured value.
The shift value is displayed on the sub display which can be
increased or decreased by
/
.
The data is shifted by the value you input. This function
may be used for fine tuning around turbidity zero.
10.00
to
10.00
0.00