Installation
17
Influence of soiled piping
If a low flow rate is set, solids may flake off from the soiled sampling piping due to the
fluctuations in the flow rate or vibration of the piping. This may eventually influence the
turbidity. In order to avoid this, minimize the influence by sampling at as constant a flow rate
as possible, or by increasing the flow rate.
Note
If the automatic cleaner is operated while the inlet/outlet valve for the detector filled with water is
closed, the automatic cleaner will be damaged. If the automatic cleaner is provided, do not seal the
detector.
Tip
It is recommended that pressurized air can be used for sampling pipe cleaning.