Measurement
43
Output type
setting for
calibration
Specify the calibration contact type.
CAL: Issues an output in a hold state during and after zero calibration
trg: Outputs the contact for 5 seconds starting 2 seconds after the
completion of calibration
CAL.E: Issues an output when a calibration error occurs
However, this error does not maintain the transmission output in a hold
state.
Note:
This is displayed when “1.tyPE” is set to “CAL”.
CAL, trg,
CAL.E
Error alarm (FAiL) output setting: RF
Setting of FAiL
output when
abnormality is
detected during
self-check
Specify whether or not to output a FAiL when an abnormality is detected
during self-check.
(E-80 to 85)
yES, no
Setting of FAiL
output when
measured
value exceeds
full-scale range
Specify whether or not to output a FAiL when the measured value
exceeds the full-scale range.
Over full scale is regarded as “FAiL”.
When the FAiL is set to be output, it is output if the measured value
exceeds the following value:
When the range cut is enabled:
PSL: 100.0 degree, Formazin: 2000 degree, Kaolin: 1000 degree
When the range cut is disabled:
PSL: 110.0 degree, Formazin: 2200 degree, Kaolin: 1100 degree
yES, no
Setting of FAiL
output when a
calibration
error occurs
(Calibration
error is
regarded as
“FAiL”)
Specify “yES” when a FAiL is to be output if a calibration error occurs. In
this case, a transmission output is held in the meas mode until the FAiL is
cancelled.
yES, no
FAiL output
delay time
setting
Specify the time in seconds from when a FAiL occurs until a contact is
output.
Note:
If a failure is restored, FAiL contact will be cancelled without delay.
0 to 600
(s)
General transmission output setting
Transmission
hold plan in
general
Specify the transmission holding status in general in the output hold
mode.
HoLd: The current will be held at the latest value
PrES: The current will be held at the preset value
HoLd,
PrES
Display
Item
Description
Setting
range