Measurement
44
Burnout hold
setting for
general
transmission
output
Specify the burnout current of the transmission output when an error
occurs during measurement.
out.4: Outputs 3.8 mA at the time of burnout
out.20: Outputs 21 mA at the time of burnout
non: No burnout setting
Note:
A burnout current is output in the meas mode only.
non,
out.4,
out.20
Transmission
output 2 range
(fixed/
automatic
changeover)
setting
For transmission output 2, a maximum of four ranges can be switched
using the external contact. In addition, the converter selects the optimal
range according to the automatic range changeover function.
mg: Sets the range using the external input (sets the range using
EXT3, 4)
Auto: Uses the automatic range changeover function
rng, Auto
Transmission output 1 setting
Transmission
output 1 full-
scale range
(fixed range)
setting
Select the transmission output range. The alarm value can be specified
based on the selected decimal point.
10.00,
100.0,
1000
Zero point
value setting
for
transmission
output 1
Specify the value at the zero point (4 mA) of transmission output 1.
Any value within the range can be set.
Note:
The item of sub display depends on which ever “10.00”
*1
, “100.0”
*2
or “1000”
*3
in “C.tgt” of Cur1 is selected.
*1
0.00 to
11.00
*2
0.0 to
110.0
*3
0 to 2200
Span point
value setting
for
transmission
output 1
Specify the value at the span point (20 mA) of transmission output 1.
Any value within the range can be set.
Note:
The item of sub display depends on which ever “10.00”
*1
, “100.0”
*2
or “1000”
*3
in “C.tgt” of Cur1 is selected.
*1
0.00 to
11.00
*2
0.0 to
110.0
*3
0 to 2200
Preset value
setting for
transmission
output 1
Input the preset value for current holding.
Note:
This is displayed when “C.HoLd” is set to “PrES” and the item of sub
display depends on which ever “10.00”
*1
, “100.0”
*2
or “1000”
*3
in
“C.rng” of Cur1 is selected.
*1
0.00 to
11.00
*2
0.0 to
110.0
*3
0 to 2200
Transmission output 2.A to 2.d setting
Tip
The other transmission output 2.b, 2.c, 2.d can be set in the same manner.
The transmission output 2 provides maximum four output ranges which are selected by
external inputs or automatic range selection function. The zero and span point of the
transmission range can be set any within the measuring range.
Transmission
output 2.A
value (full-
scale range)
setting
Select the transmission output range. The alarm value can be specified
based on the selected decimal point.
10.00,
100.0,
1000
Display
Item
Description
Setting
range