Functional Tests
8500C/8500C+ System Maintenance Manual
5-5
NOTE: To undo mistaken entries, press the RESTORE soft key.
h.
On same line, press right arrow key twice to select last number (called “times factor”) with
arrow keys.
i.
Key in 2 as new times factor. This entry now reads: x 2.
j.
Select
BALANCE TYPE
field with arrow keys.
k.
Press
CHANGE
soft key to select <
MAG PICKUP
>.
l.
Select <
AUTO STOP
> and press
CHANGE
soft key to select <
MAN STOP
>.
m.
Select
PHASE UNITS
field with arrow keys.
n.
Press
CHANGE
soft key to select
<DEGREES>
.
o.
Select
PICKUP
field with arrow keys.
p.
Press
CHANGE
soft key to select
CHANNEL A
.
q.
Select
INPUT
field with arrow keys.
r.
Key in
<1>
with numeric key.
s.
Press
ENTER
soft key.
t.
Wait 10 seconds.
u.
Press
START
key.
v.
Verify that display indicates (example in Figure 5-3):
900 +1 RPM
0.4 +0.04 IPS
275° +10°
ACQUISTION DISPLAY
FINAL RESULTS DISPLAY
Figure 5-3. Typical Balance Test Displays for 900 RPM