Functional Tests
8500C/8500C+ System Maintenance Manual
D-9
p.
Select
PICKUP
field with arrow keys.
q.
Press
CHANGE
soft key to select
PICKUP C
.
r.
Select
INPUT
field with arrow keys.
s.
Key in
3
with numeric key.
t.
Press
ENTER
soft key.
u.
Wait 10 seconds.
v.
Press
START
key.
w.
Verify that final results display indicates.
1800 +2 RPM
0.8 +0.10 IPS
NOTE: This response assumes that internal charge converters have been properly
programmed for the sensors under test. (ref. Section D.6 for charge amplifier programming/
calibration).
x.
Press
STOP
key.
y.
Press
SETUP
key.
z.
Select
INPUT
field with arrow keys.
aa.
Key in
6
with numeric key.
bb.
Press
ENTER
soft key.
cc.
Wait 10 seconds.
dd.
Press
START
key.
ee.
Verify that final results display indicates.
1800 +2 RPM
0.8 +0.10 IPS
ff.
Press
STOP
key.
gg.
Move 10700 cables to High Temp Accelerometer 9 and 12 channel positions and move
cable 14883 to photocell channel D.
hh.
Press
SETUP
on 8500C/C+.
ii.
Select
PICKUP
field with arrow keys.
jj.
Press
CHANGE
soft key to select pickup shown in Table D-3.
kk.
Select
INPUT
field with arrow keys.
ll.
Key in channel number 9 with numeric key.
mm.
Press
ENTER
soft key.
nn.
Wait 10 seconds.