D-10
Chadwick-Helmuth A Business Unit of Honeywell International, Inc.
Appendix D -
8520CS Signal Selector
oo.
Press
START
key.
pp.
Verify that final results display indicates.
1800 +1 RPM
0.8 +0.10 IPS
qq.
Press
STOP
key.
rr.
Repeat steps
gg.
through
pp
. for channel 12.
D.5.3
FasTrak Test
D.5.3.1 Test Setup (Ref. Figure D-5)
NOTE: This test provides a high level of confidence that the FasTrak is operating properly.
Before doing the FasTrak test, apply the special overlay to the calibrator disk. See Figure D-6 for
samples that can be copied or recreated. Unplug the 14883 cable from the “D” connector on the
8520CS Signal Selector.
NOTE: This test should be performed in a well-lit room.
a.
Place calibrator on its back on a flat surface.
b.
Place FasTrak on its side with the pin connector facing to the right. The window should be
facing the calibrator approximately six inches away.
c.
Set
CAM RATE (RPM)
switch to
900
.
d.
Set
MOTOR
switch to
ON
.
e.
Adjust pot next to
900 CAM RATE
switch position until 2nd ring from outside on rotor
disk appears frozen in place under 60 Hz light.
2
f.
Press
BALANCE
key on 8500C/C+.
g.
Press
SETUP
key.
h.
Select
PICKUP
field with arrow keys.
i.
Press
CHANGE
soft key to select
PICKUP B
.
2
Attention, users outside U.S.: The 11A Calibrator can also be calibrated under 50Hz light. For 1714
rpm and 923 rpm cam rates, inner two rings on rotor disk are used. Remember to enter half these rates
in 8500C/C+ setup screen, when called for in the following steps.
Table D-3. Accelerometer and Photocell Balance Test Channel Combinations
Accelerometer Channels
Photocell Channels Channel
3 and 6
C
9 and 12
D