3.2 Automatic Continuous Measurement (6-cell auto mode)
3 - 28
Table 3-9 Auto Zero Interval and Operation of Instrument (1)
Sequence of Operation
Auto Zero Interval: 5
Auto Zero Interval: 1
Cell Position
Operation
Setting of standards (1st round)
1
1
Cell A
Auto zero
2
2
Cell A
STD1 measurement
⎯
3
Cell A
Auto zero
3
4
Cell 2
STD2 measurement
⎯
5
Cell A
Auto zero
4
6
Cell 3
STD3 measurement
⎯
7
Cell A
Auto zero
5
8
Cell 4
STD4 measurement
⎯
9
Cell A
Auto zero
6
10
Cell 5
STD5 measurement
Setting of standards (2nd round)
7
11
Cell A
Auto zero
8
12
Cell 1
STD6 measurement
⎯
13
Cell A
Auto zero
9
14
Cell 2
STD7 measurement
Repeated until the specified number of standards are all measured.
Table 3-10 How to Set Cells with Blank Selected for
STD Autozero
Cell A
Cell 1
Cell 2
Cell 3
Cell 4
Cell 5
1st
round Blank STD1 STD2 STD3 STD4 STD5
2nd
round
Blank STD6 STD7 STD8 STD9 STD10
3rd round
Blank
STD11
STD12
STD13 STD14 STD15
4th round
Blank
STD16
STD17
STD18 STD19 STD20