3.3 Sample-by-sample Measurement (6-cell manual mode)
Photometry
CELL A
1 Meas.PARAM.
6 Cell Mode
Auto
2 Sample
SAMP.Autozero ON
3 6 Cell
Autozero int.
5
4 Print
No.of Sample
1
5 Save Method
6 Meas.Screen
Select function. : _
□
1 Auto □
2 Manual
Fig. 3-79 6-cell Condition (6 Cell) Screen
(3) To set the 6-cell conditions, press the [ENTER] key.
Or select each item with the [
▲
] or [
▼
] key and make
your setting according to the guidance. For details of
each parameter, refer to Table 3-48.
(4) After selecting “Manual” for 6 Cell Mode, return to the
Photometry screen by pressing the [RETURN] key or
[
◄
] key .
Table 3-48 “6 Cell” Setting Parameter
Setting Item
Description
6 Cell Mode
Specify a 6-cell operation mode for
measurement.
1) Auto
In this mode, a sample is set at each
position of the 6-cell holder beforehand
and the holder is automatically turned to
automatically perform from auto zero to
standard/sample measurement. It is
recommended to use this mode if you
have 6 cells and there are many samples
to be measured. (For details of the Auto
mode, refer to section 3.3.3..)
2) Manual
In this mode, standard/sample
measurement is carried out by using one
position (cell 1 to 5 selectable) of the 6-cell
holder. To eliminate the influence of
instrument drift (baseline fluctuation)
during measurement, auto zero can easily
be executed by setting a solution for auto
zero to cell A. In normal usage, set a
standard or sample solution to cell 1.