3.2 Automatic Continuous Measurement (6-cell auto mode)
3 - 42
(3) The Photometry screen (Fig. 3-23) appears.
Photometry
CELL A
1 Meas.PARAM
Data Mode CONC
2 Sample
No.of WL 1
3 Curve
WL1(nm)
600.0
4 Curve Data
Delay(s) 0
5 6 Cell
6 System
7 Print
8 Save Method
9 Meas.Screen
Select function. : _
Fig. 3-23 Photometry Screen
(4) To set the measuring conditions, press the <1> key
(Meas. PARAM.) and then [ENTER] key. Or select
“Meas. PARAM.” with the [
▲
] or [
▼
] key and press the
[ENTER] key or [
►
] key.
(5) The measuring condition (Meas. PARAM.) screen (Fig.
3-24) appears.
Photometry
CELL A
^
1 Meas.PARAM.
Data Mode
CONC
2 Sample No.of WL
1
3 Curve WL1(nm)
600.0
4 Curve Data Delay(s)
0
5 6 Cell
6 System
^7 Print
8 Save Method
9 Meas.Screen
Select function. : _
□
1
CONC □
2
ABS □
3
%T □
4
DNA
Fig. 3-24 Measuring Condition (Meas. PARAM.)
Screen
(6) For Data Mode, press the <2> key (ABS) for
absorbance measurement or <3> key (%T) for
transmittance measurement and then press the
[ENTER] key.
(7) With the [
▲
] or [
▼
] key, set the number of
wavelengths (No. of WL) and wavelength (WL).
For details of each parameter, refer to Table 3-19.