3.2.1
3 - 31
(6) When the specified number of standards have all been
measured, the screen shown in Fig. 3-16 appears.
When you select the <1> key (Sample Meas.), the
system terminates standard measurement and
proceeds to sample measurement. Selecting the <2>
key (Curve), <3> key (Save) and <4> key (Print) allows
the processings shown in Table 3-13.
Photometry
CELL A
ID ABS CONC
STD6 0.498 50.00
STD7 0.602 60.00
STD8 0.707 70.00
STD9 0.800 80.00
STD10 0.893 90.00
Select function. : _
□
1
Sample Meas. □
2
Curve □
3
Save □
4
Fig. 3-16 Guidance after Standard Measurement
(No. of STD: 10)
Table 3-13 Guidance after Standard Measurement
Setting Item
Description
[1] Sample
Meas.
The system proceeds to sample measurement.
[2] Curve
You can generate a calibration curve with the
measured standards and confirm it. When this
function is selected, the following screen appears.
Photometry/Curve
CELL A
R=0.9999 R2=0.9999
Select function. : _
□
1
Scale □
2
Factor □
3
STD
ABS
0.000
90.00
CONC
0.893
0.001
+
+
+
+
+
+
+
+
+