4.3.3 Deletion and Recovery of Calibration Curve Data
1.
Displaying Calibration Curve Data
[For check from saved data]
For check by opening the saved data, open the photometric
data including the saved calibration curve data according to
section 4.1.1.
Upon opening the calibration curve data, a guidance (an
example shown in Fig. 4-29) is presented. Select the <2>
key (for Curve).
Select function. :_
[1]Meas. [2]Curve [3]Save [4]Print
◄
►
( 1/ 1)
Fig. 4-32 Example of Guidance upon
Opening of Calibration Curve Data
[During measurement]
During measurement by selecting Photometry in the Main
Menu, Conc for Data Mode and 1st Order for Curve Type,
select Curve Confirmation or Curve indicated in the
guidance.
2.
Displaying Standard Sample Data
(1) The calibration curve data is now opened. Select the
<3> key (for STD).
Photometry/Curve
CELL A
R = 0.9999 R2 = 0.9999
ABS
0.000 90.000
CONC
0.001
+
+
+
+
+
+
+
+
+
0.893
Select function. :_
[1]Scale [2]Factor [3]STD
Fig. 4-33 Calibration Curve Data
4 - 29