4.3 Data Processing
Table 4-16 Scale Changing Parameters for Time Scan
Function Key
Explanation
X-Max
Sets the maximum value on the X axis (time axis) of
spectrum.
Settable within 10 to 99999. For adopting and
changing the set axis, return to the spectrum screen
by pressing the [RETURN] key.
X-Min
Sets the minimum value on the X axis (time axis) of
spectrum.
Settable within 0 to 99989. For adopting and
changing the set axis, return to the spectrum screen
by pressing the [RETURN] key.
Y-Max
Sets the maximum value on the Y axis (representing
the absorbance) of calibration curve.
Settable range is as follows.
Absorbance : -9.999 to 9.999
Transmittance : -9.999 to 999.9
For adopting and changing the set axis, return to the
spectrum screen by pressing the [RETURN] key.
Y-Min
Sets the minimum value on the Y axis (representing
the absorbance) of spectrum.
Settable range is as follows.
Absorbance : -9.999 to 9.999
Transmittance : -9.999 to 999.9
For adopting and changing the set axis, return to the
spectrum screen by pressing the [RETURN] key.
Scale Reset
Sets scale in the default value. For setting, press
the <1> key (Execute).
Select function. :_
[1]Execute
Auto Scale
Judges the maximum and minimum values of the
measured data on the Y axis, and automatically
optimizes the scale. For setting, press the <1> key
(Execute).
Select function. :_
[1]Execute
4 - 34