3.2 Automatic Continuous Measurement (6-cell auto mode)
3 - 78
(cont’d)
Setting Item
Description
[1] Meas.
Photometry
CELL A
ID ABS
10 WL1 ( 260.0nm) 0.423
WL2 ( 280.0nm) 0.268
ABS DIFF= 0.155
ABS Ratio= 1.582
Set sample to Cell1(Sample10).: __
START:Meas.; AUTOZERO:Autozero
W
X
( 10/ 10)
[2] Save
You can save the measured data. When this
function is selected, the following screen appears.
Save Data
CELL A
1 NO3-(Hitachi Taro) PHT 2010/10/10
2 Cr6+(Hitachi Hanako)
PHT 2010/03/30
Select function. : _
□
1
Save □
2
Sort
W
X
( 1/ 1)
[1] Save : Allows you to save data under a new file
name.
[2] Sort : Allows you to rearrange the currently
saved data in the order of file names or
dates.
[3] Print
You can select items to be printed. When this
function is selected, the screen given below appears.
For each item, select whether to print out (ON or
OFF).
Photometry
CELL A
Print Date
ON
Run Date
ON
Method
ON
Sample
ON
Select function. : _
□
1
ON □
2
OFF