3.3 Sample-by-sample Measurement (6-cell manual mode)
4.
Setting of System Conditions
(1) To set the system conditions, press the <3> key
(System) and then [ENTER] key. Or select “System”
with the [
▲
] or [
▼
] key and press the [ENTER] key or
[
►
] key.
(2) The system condition (System) screen (Fig. 3-94)
appears.
Time Scan
CELL A
1 Meas.PARAM.
Response
Medium
2 Sample
3 System
4 Print
5 Save Method
6 Meas.Screen
Select function. : _
□
1 Fast □
2 Medium □
3 Slow
Fig. 3-94 System Condition (System) Screen
(3) Set up a response. After completion of setting, return
to the Time Scan screen by pressing the [RETURN]
key or [
◄
] key.
Setting Item
Description
Response
Select from the following 3 response speeds.
[1] Fast
: Used for high-resolution
measurement with respect
to the wavelength.
Note that the noise level is
high as compared with
Medium and Slow.
[2] Medium : Used for ordinary
measurement.
[3] Slow
: Used to reduce a variation
in photometric value.
This response speed is
unsuitable for sample
measurement reflecting an
abrupt change with time as
compared with Fast and
Medium.
3 - 138