3.2.3
3 - 75
<1> Press the [AUTOZERO] key when the sample setting screen is
displayed.
Photometry
CELL A
[Info]
Set samples to Cell No..
Cell A
-
Cell 1
Sample1
Cell 2
Sample2
Cell 3
Sample3
Cell 4
Sample4
Cell 5
Sample5
START:Meas.:_
Sample setting screen
<2> The auto zero execution screen appears. Set an auto zero
sample to cell A. The lamp is lit for 30 seconds and the current
measured value is indicated so that you can check if execution
of auto zero is necessary.
Photometry 600.0nm 0.001ABS
CELL A
[Info]
Set autozero sample to Cell A.
AUTOZERO:Execute Autozero; CLEAR:Cancel
Auto zero execution screen
Fig. 3-44 Manual Auto Zero with OFF set for
SAMP. Autozero
Upon completion of sample setting according to the
guidance, press the [START] key. Measurement will
then start. Measurement is carried out for every 5
samples. Guidance is given until the specified
number of samples are all measured.
GUIDE
: For how to set a sample to the cell holder, refer
to section 2.4.4.