3.2 Automatic Continuous Measurement (6-cell auto mode)
3 - 64
Explanation 3-8 Background Correction
Bkgd.CORR Setting
method
OFF
Background correction is avoided. Absorbance ratio will be calculated using the
obtained absorbance values.
Absorbance ratio = A(1)/A(2)
WL2 WL1
Wavelength
Absorption spectrum
ON
The background will be corrected. This is effective when the background level is
even along the wavelength axis. The absorbance at the set correction wavelength
will be subtracted from those at the measuring wavelengths. Calculation is carried
out according to the following equation.
Absorbance ratio = (A(1) - A(correction))(A(2) - A(correction))
WL2 WL1
Correction
wavelength
Wavelength
Absorption spectrum
Absor
bance
A(2)
A(1)
Absorbance
A(2)
A(1)
A (correction)