Appendix 12 Instrument/Unit Communications Interruptions and Reestablishment
A
36
The instrument, measurement units, and wireless loggers are synchronized as they com-
municate during measurement. Consequently, since time will progress at each device's
own level of precision in the event that communications with measurement units or wire-
less loggers are interrupted during measurement, time differences will appear among the
three devices. In some cases, there may be discrepancies such as those illustrated below
in the number of data points captured once communications between the instrument and
the measurement units and wireless loggers are reestablished following an interruption. In
the event of discrepancies in the number of data points or time between the instrument on
the one hand and the measurement units and wireless loggers on the other, data will be
restored based on the assumption that the instrument's data is correct.
Synchronization and acquired data discrepancies
IMPORTANT
Data continuity may be compromised when a discrepancy occurs.
Summary of Contents for LR8410-20