![Hioki LR8410-20 Instruction Manual Download Page 141](http://html1.mh-extra.com/html/hioki/lr8410-20/lr8410-20_instruction-manual_2137134141.webp)
5.1 Triggering Measurement Start and Stop
129
3
5
Chapter 5 S
p
ecif
ying Crite
ria
for M
easurement
s
There are two ways to set the criteria to start and stop recording according to wave-
form slope.
Select whether to trigger at the rising or falling edge of the input signal.
5.1 Triggering Measurement Start and Stop
Types of Trigger Criteria
Type of Analog
Trigger
Trigger Example
Description
Level Trigger
[Level]
Recording starts (or stops) when the
input signal crosses the specified trig-
ger level (voltage).
Window Trigger
[In]
Recording starts (or stops) when the
input signal enters a range defined by
upper and lower thresholds.
[Out]
Recording starts (or stops) when the
input signal exits a range defined by
upper and lower thresholds.
Pattern Trigger
[1]
Measuring starts and stops when
the input signal logic switches
from Low to High.
[0]
Measuring starts and stops when
the input signal logic switches
from High to Low.
[X]
Signal is ignored. No triggering
occurs.
Trigger Level
Input
Trigger Slope
[
]
[
]
T
T
Upper
Threshold
Lower
Threshold
T
T
Upper
Threshold
Lower
Threshold
T
T
T T
T
HIGH
LOW
T T
HIGHHigh
LOWLow
High
Low
Summary of Contents for LR8410-20
Page 1: ......
Page 2: ......
Page 12: ...Contents x...
Page 32: ...Operating Precautions 20...
Page 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Page 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Page 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Page 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Page 316: ...13 3 Disposing of the instrument 304...
Page 360: ...Index Index 4...
Page 361: ......
Page 362: ......
Page 363: ......
Page 364: ......