Appendix 2 Error Messages and Remedial Actions
A
4
Appears just once when an error occurs. Disappears within a few seconds. Also disap-
pears when any key is pressed.
Warning Messages
Message
Remedial Action
101
Invalid key.
The pressed key is disabled due to measuring or
other operation. Press the key again after the cur-
rent operation finishes.
103
Cannot change while measuring.
Press
STOP
to stop measuring before changing the
setting.
104
No measurement data was found
for the specified position.
Specify the location at which measurement data ex-
ists in the internal buffer memory, or save the mea-
surement data to a storage medium and then load
the data from a specified location into the internal
buffer memory.
201
Exceeding the setting range.
The entered numerical value is out of the valid
range. Enter a valid value.
204
Measurement range changed.
When
[Type]
-
[Up/Lwr]
is selected, the measure-
ment range is automatically changed to match the
upper and lower limits.
230
Recording time has been changed
due to memory limit.
Increasing the number of measurement channels
has decreased Internal memory waveform
data capacity.
231
Pre-Trigger span changed.
Increasing the recording interval or length requires
reducing pre-trigger recording time.
232
Recording and Pre-Trigger spans
changed due to waveform
memory limit.
Increasing the recording interval requires reducing
recording length and pre-trigger recording time.
234
Burn Out (broken-wire) detection
changed.
Burn-out detection is not available with the current
settings. To enable burn-out detection, select a lon-
ger recording interval or reduce the number of chan-
nels used.
See:
"About the Recording Interval" (p .70)
235
Burnout detection cannot be
turned On for the current
recording interval setting.
Change the recording interval to a
value from 200 ms to 1h.
The setting is not available due to the combination of
the number of channels used and the recording in-
terval. To enable burn-out detection, select a longer
recording interval, or reduce the number of channels
used.
See:
"About the Recording Interval" (p .70)
236
File partition period was changed.
The segment time span has been changed in re-
sponse to changing the recording interval.
237
File partition time was changed.
The segment time span has been changed in re-
sponse to changing the recording interval.
238
Unable to set CSV auto-save with
the current record interval.
The recording intervals available for CSV auto sav-
ing are limited. Select a longer recording interval.
See:
"Automatic Saving (Waveform Data and
Numerical Calculation Results)" (p .163)
2
Summary of Contents for LR8410-20
Page 1: ......
Page 2: ......
Page 12: ...Contents x...
Page 32: ...Operating Precautions 20...
Page 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Page 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Page 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Page 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Page 316: ...13 3 Disposing of the instrument 304...
Page 360: ...Index Index 4...
Page 361: ......
Page 362: ......
Page 363: ......
Page 364: ......