Appendix 10 Introduction to Measurement Applications
A
29
Appendix
This section describes how to measure current fluctuations by connecting an instrumenta-
tion device that generates 4 to 20 mA output. We will record average values every minute.
Appendix 10Introduction to Measurement Appli-
cations
Connecting a 4 to 20 mA Output Device and Recording Average
Values Every Minute
1
Prepare the Following Before Measuring
Items to prepare
LR8410-20
Measurement unit
AC Adapter (supplied)
Measurement (input)
leads
250
shunt resistor
SD memory card
*:Included or optional
1
Place the instrument and measurement unit close
together.
2
Connect them to a wall outlet.
3
Turn on both the instrument and the measurement
unit.
4
Register the measurement unit with the instrument.
5
Insert the SD memory card (exercising care to orient
it properly).
6
Install the measurement unit in the measurement
location (verify that the UNIT LED for the registered
unit is lit up).
7
Connect the input cable and the 250
shunt resis-
tor to CH1 (p. 47).
8
Connect the instrumentation device to the output
terminal.
Summary of Contents for LR8410-20
Page 1: ......
Page 2: ......
Page 12: ...Contents x...
Page 32: ...Operating Precautions 20...
Page 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Page 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Page 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Page 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Page 316: ...13 3 Disposing of the instrument 304...
Page 360: ...Index Index 4...
Page 361: ......
Page 362: ......
Page 363: ......
Page 364: ......