Appendix 8 Concerning Noise Countermeasures
A
17
Appendix
1st. The Importance of Device Grounding
Ground the measurement units securely
The measurement units is designed so that the chassis is grounded when the three-
prong power cord from the adapter is plugged directly into a grounded, three-prong
receptacle.
Securely ground the chassis of the device being tested
Securely connect the chassis of the device being tested to a good ground.
The wireless logger's design does not permit it to be grounded. If the unit is being
affected by noise, operate it using battery power.
Noise Countermeasure Example
Thermo-
couple
Device being Tested
measurement
unit
Grounding Both the Device Being
Tested and the measurement unit