![Hioki LR8410-20 Instruction Manual Download Page 298](http://html1.mh-extra.com/html/hioki/lr8410-20/lr8410-20_instruction-manual_2137134298.webp)
12.2 LR8510 Wireless Voltage/Temp Unit Specifications
286
12.2 LR8510 Wireless Voltage/Temp Unit Specifi-
cations
Basic Specifications
Product warranty
period
3 years
AC adapter: 1 year
Connector: Not covered by the warranty
Period of guaranteed
accuracy
1 year
Supported model
LR8410-20 Wireless Logging Station
Control and
communications
(Between instrument
and unit)
Bluetooth 2.1
+
EDR
Communications range: 30 m, line of sight
Security: SSP
Backup memory
When recording n channels: (65,536/n) data points
Data is maintained in the event of a communications error and resent
when communications are restored.
Operating environment
Indoors, Pollution degree 2, up to 2000 m (6562-ft.) ASL
Operating temperature
and humidity
Temperature: -20
C to 60
C(-4°F to 140°F)
Humidity: -20
C to 40
C(-4°F to 104°F)80%RH or less (noncondensating)
40
C to 45
C(104°F to 113°F)60%RH or less (noncondensating)
45
C to 50
C(113°F to 122°F)50%RH or less (noncondensating)
50
C to 60
C(122°F to 140°F)30%RH or less (noncondensating)
(temperature variation range is 5 to 35°C (41 to 95°F))
Storage temperature
and humidity
Temperature: -20
C to 60
C(-4 to 140°F)
Humidity: -20
C to 40
C(-4°F to 104°F)80%RH or less (noncondensating)
40
C to 45
C(104°F to 113°F)60%RH or less (noncondensating)
45
C to 50
C(113°F to 122°F)50%RH or less (noncondensating)
50
C to 60
C(122°F to 140°F)30%RH or less (noncondensating)
Conditions of
guaranteed accuracy
Defined after zero-adjustment has been performed.
The 50/60 Hz cut-off setting is selected
(see the Cut-Off Frequency Table)
Temperature and
humidity range for
guaranteed accuracy
Temperature: 23±5°C (73±9°F), Humidity 80% RH or less (non-conden-
sating)
Dielectric strength
AC3.0 kV/60 seconds (between each input channel and chassis)
AC350 V/60 seconds (between input channels)
Dimensions
Approx.150W×90H×56D mm (5.91"W × 3.54"H × 2.2"D)
(including cover)
Mass
Approx. 340 g (12.0 oz.)
Applicable Standards
Safety: EN61010
EMC:
EN61326 Class A
Summary of Contents for LR8410-20
Page 1: ......
Page 2: ......
Page 12: ...Contents x...
Page 32: ...Operating Precautions 20...
Page 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Page 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Page 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Page 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Page 316: ...13 3 Disposing of the instrument 304...
Page 360: ...Index Index 4...
Page 361: ......
Page 362: ......
Page 363: ......
Page 364: ......