12.2 LR8510 Wireless Voltage/Temp Unit Specifications
290
Thermocouple
burnout detection
Enable/disable thermocouple burn-out detection at each recording inter-
val.5
A±20% detection current.
No current flow when acquiring measurement data.
(The burnout detection setting cannot be used with a recording interval
of 100 ms.)
Digital filter
Off, 50 Hz, 60 Hz (automatically sets the cut-off frequency according to
the recording interval)
Cut-off frequency
Input resistance
1 M
±5% (voltage and thermocouple measurement)
Normal mode rejection
ratio
50dB minimum (with 50 Hz input with 5 s recording interval, and 50 Hz
digital filter enabled)
(with 60 Hz input with 5 s recording interval, and 60 Hz digital filter
enabled)
Common mode
rejection ratio
with 100
maximum signal source impedance
100dB minimum (with 50 Hz or 60 Hz input and digital filter disabled)
140dB minimum
(for 50-Hz input with 5 s recording interval, 50-Hz digital filter enabled,
10 mV f.s. range setting)
(for 60-Hz input with 5 s recording interval, 60-Hz digital filter enabled,
10 mV f.s. range setting)
Maximum input voltage
DC±100 V
Max. inter-channel
voltage
DC300 V
Maximum rated voltage
to earth
300 VAC, DC (between each input channel and chassis)
Measurement category II
Anticipated transient overvoltage 2,500 V
Inputs
Digital filter
settings
Burn-
out
detec-
tion
set-
ting
Recording interval
100
ms
200
ms
500
ms
1 s
2 s
5 s
10 s
20 s
30 s to
60 Hz
OFF 1.8 k 0.7 k 60
60
60
60
60
60
60
ON
-
1.8 k 0.7 k 60
60
60
60
60
60
50 Hz
OFF 1.8 k 0.7 k 50
50
50
50
50
50
50
ON
-
1.8 k 0.7 k 50
50
50
50
50
50
OFF
OFF 1.8 k 1.8 k 1.8 k 1.8 k 1.8 k 1.8 k 1.8 k 1.8 k 1.8 k
ON
-
1.8 k 1.8 k 1.8 k 1.8 k 1.8 k 1.8 k 1.8 k 1.8 k
[Unit:Hz]
Summary of Contents for LR8410-20
Page 1: ......
Page 2: ......
Page 12: ...Contents x...
Page 32: ...Operating Precautions 20...
Page 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Page 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Page 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Page 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Page 316: ...13 3 Disposing of the instrument 304...
Page 360: ...Index Index 4...
Page 361: ......
Page 362: ......
Page 363: ......
Page 364: ......